datasheetbank_Logo
전자부품 반도체 검색엔진( 무료 PDF 다운로드 ) - 데이터시트뱅크

EDGE4707B 데이터 시트보기 (PDF) - Semtech Corporation

부품명
상세내역
제조사
EDGE4707B Datasheet PDF : 25 Pages
1 2 3 4 5 6 7 8 9 10 Next Last
Edge4707B
TEST AND MEASUREMENT PRODUCTS
Circuit Description
Circuit Overview
Control Inputs
The Edge4707B is a quad channel parametric test and
measurement unit that can :
Force Voltage / Measure Current
Force Current / Measure Voltage
Force Voltage / Measure Voltage
Force Current / Measure Current
FV/FI* is a TTL compatible input which determines whether
the PMU forces voltage or current, and MI/MV* is a TTL
compatible input which determines whether the PMU
measures current or voltage. FV/FI* and MI/MV* are
independent for each PMU. Table 1 describes the modes
of operation controlled by these pins.
Each PMU channel can force or measure voltage over a
15V range and force or measure current over four distinct
ranges:
± 2 µA
± 20 µA
± 200 µA
± 2 mA.
An on-board window comparator provides two bit output
range classification. Also, a monitor passes a real time
analog voltage which tracks either the measured current
or voltage.
PPMU Functionality
The trapezoid in Figure 1 describes the current-voltage
functionality of the PMU with VCC = 15.5V and
VEE = 4.5V, in Range D.
V
VOUT (@ I = 0) = 13.25V
No restrictions
VCC = +15.5V
VOUT (@ 200 µA) = 12.8V (in Range D)
VOUT (@ 2 mA) = 11.25V
FV / FI*
0
0
1
1
MI/MV*
0
1
0
1
Mode of Operation
Force Current, Measure Voltage
Force Current, Measure Current
Force Voltage, Measure Voltage
Force Voltage, Measure Current
Table 1.
RS0 and RS1 are TTL compatible inputs to an internal
analog mux which selects an external resistor
corresponding to a desired current range. The truth table
for RS0 to RS1, along with the associated external resistor
values and current ranges, is shown in Table 2. RS0 and
RS1 are independent for each channel of the 4707B.
RS1
RS0
Range
Current
Range
"Nominal" Ext. R
0
0
A
± 2 µA
RA = 1M
0
1
B
± 20 µA
RB = 100K
1
0
C
± 200 µA
RC = 10K
1
1
D
± 2 mA
RD = 1K
IMIN (–2 mA)
IMAX (2 mA)
FORCE/SENSE
Table 2.
VOUT (@ –200 µA) = –1.8V (in Range D)
VOUT (@ I = 0) = –2.25V
VEE = –4.5V
VOUT (@ 2 mA) = –0.25V
NOTE: Negative current implies current is flowing into the 4707 from DUT.
Figure 1. PMU Functionality
Revision 3 / December 18, 2002
6
FORCE is an analog output which either forces a current
or forces a voltage, depending on which operating mode
is selected.
SENSE is a high impedance analog input which measures
the DUT voltage input in the MV operating mode.
FORCE and SENSE are brought out to separate pins to
allow remote sensing.