datasheetbank_Logo
전자부품 반도체 검색엔진( 무료 PDF 다운로드 ) - 데이터시트뱅크

H7422P-50 데이터 시트보기 (PDF) - Hamamatsu Photonics

부품명
상세내역
제조사
H7422P-50
Hamamatsu
Hamamatsu Photonics 
H7422P-50 Datasheet PDF : 4 Pages
1 2 3 4
Metal Package PMT with Cooler
Photosensor Modules H7422 Series
Heatsink with fan (A7423) sold separately
The H7422 series are photosensor modules with an internal high-voltage power
supply circuit and a cooler installed to the metal package photomultiplier tube. Effi-
cient cooling was achieved by placing the cooler near the photomultiplier tube to
reduce thermal noise emitted from the photocathode and a high S/N ratio can be
obtained even at extremely low light levels.
The H7422-40 has high sensitivity in the 300 nm to 720 nm wavelengths. The
H7422-50 is sensitive along a wide spectral range from 380 nm to 890 nm. The
H7422-01, H7422-02 and H7422-20 have a maximum output current value of 100
µA and so are extremely effective when measurements are needed over a wide
dynamic range. The photomultiplier tube is maintained at a constant temperature
by monitoring the output from a thermistor installed near the photomultiplier and
then regulating the current to the thermoelectric cooler.
Product Variations
Type No.
H7422-40
H7422P-40
H7422-50
H7422P-50
H7422-01
H7422-02
H7422-20
Spectral Response Max. Output Signal Current
300 nm to 720 nm
380 nm to 890 nm
2 µA
300 nm to 850 nm
300 nm to 870 nm
300 nm to 890 nm
100 µA
Features
GaAsP photocathode, QE 40 % at peak
wavelength, high gain (P type)
For photon counting
GaAs photocathode, QE 12 % at peak
wavelength, high gain (P type)
For photon counting
Multialkali photocathode
Infrared-extended multialkali photocathode
Infrared-extended high-sensitivity multialkali photocathode
Specifications
Parameter
H7422 Series
Suffix
-40
-50
-01
-02
-20
Input Voltage
+11.5 to +15.5
Max. Input Voltage for Main Unit
+18
Max. Input Current for Main Unit
62
30
Max. Input Voltage for Thermoelectric Cooler
2.6
Max. Input Current for Thermoelectric Cooler
2.2
Max. Output Signal Current
2
100
Max. Control Voltage
+0.9 (Input impedance 100 k)
Recommended Control Voltage Adjustment Range
+0.5 to +0.8
+0.25 to +0.8
Effective Area
5
7
Sensitivity Adjustment Range
1: 50
1: 104
Peak Sensitivity Wavelength
580
800
400
500
630
420 nm
108
15
56
40
40
Radiant Sensitivity
550 nm
176
50
36
56
72
800 nm
90
1.2
6.4
46
Radiant Sensitivity *1 550 nm 8.8 × 104
2.5 × 104
1.8 × 104
2.8 × 104
3.6 × 104
Dark Current *1 *2 Typ.
0.4
Max.
1.0
0.5
0.03
0.08
0.1
1.3
0.08
0.2
0.25
Radiant Sensitivity *3 550 nm 1.8 × 105
5.0 × 104
Dark Count *2 *3
Typ.
100
125
Max.
300
375
Rise Time *1
1.00
0.78
Ripple Noise *1 *4 (peak to peak) Max.
0.6
Settling Time *5
0.2
Operating Ambient Temperature
+5 to +35
Storage Temperature
-20 to +50
Weight
Approx. 400
*1: Control voltage = +0.8 V PMT setting temperature 0 °C, used with C8137-02 and A7432 *2: After 30 minute storage in darkness
*3: Plateau voltage, PMT setting temperature 0 °C, used with C8137-02 and A7423
*4: Cable RG-174/U, Cable length 450 mm, Load resistance = 1 MΩ, Load capacitance = 22 pF
18*5: The time required for the output to reach a stable level following a change in the control voltage from +1.0 V to +0.5 V.
Unit
V
V
mA
V
A
µA
V
V
mm
nm
mA/W
A/W
nA
A/W
s-1
ns
mV
s
°C
°C
g

Share Link: 