Si7013-A10
4.6. Bake/Hydrate Procedure
After exposure to extremes of temperature and/or humidity for prolonged periods, the polymer sensor film can
become either very dry or very wet; in each case the result is either high or low relative humidity readings. Under
normal operating conditions, the induced error will diminish over time. From a very dry condition, such as after
shipment and soldering, the error will diminish over a few days at typical controlled ambient conditions, e.g., 48
hours of 45 %RH 55. However, from a very wet condition, recovery may take significantly longer. To accelerate
recovery from a wet condition, a bake and hydrate cycle can be implemented. This operation consists of the
following steps:
Baking the sensor at 125 °C for 12 hours
Hydration at 30 °C in 75% RH for 10 hours
Following this cycle, the sensor will return to normal operation in typical ambient conditions after a few days.
4.7. Long Term Drift/Aging
Over long periods of time, the sensor readings may drift due to aging of the device. Standard accelerated life
testing of the Si7013 has resulted in the specifications for long-term drift shown in Table 4 and Table 5. This
contribution to the overall sensor accuracy accounts only for the long-term aging of the device in an otherwise
benign operating environment and does not include the effects of damage, contamination, or exposure to extreme
environmental conditions.
20
Rev. 1.0